Remote Control Commands
R&S
®
SMC100A
264
Operating Manual 1411.4060.32 ─ 04
Parameters:
<Frequency>
float
Increment: 0.1 Hz
*RST:
1KHz
Example:
LFO2:FREQ 5kHz
sets the frequency of the LF generator 2 signal to 5 kHz.
[:SOURce]:LFOutput:FREQuency:MANual
<Manual>
In Sweep mode (
SOUR:LFO:FREQ:MODE SWE
) the command sets the frequency for the
next sweep step in the "Step" sweep mode (
SOUR:LFO:SWE:MODE MAN
). Here only
frequency values between the settings
SOUR:LFO:FREQ:STAR
and
...:STOP
are per-
mitted. Each sweep step is triggered by a separate
SOUR:LFO:FREQ:MAN
command.
Parameters:
<Manual>
float
Range:
STARt to STOP
Increment: 0.1 Hz
*RST:
1 kHz
Example:
LFO:SWE:MODE MAN
sets the "Step" sweep mode.
LFO:FREQ:MAN 5 kHz
sets an LF frequency of 5 kHz for the next step in the "Step" sweep
mode.
LFO:FREQ:MODE SWE
sets the LF Sweep mode. An LF frequency of 5 kHz is output.
LFO:FREQ:MAN 5.1 kHz
triggers the next sweep step with a frequency of 5.1 kHz.
[:SOURce]:LFOutput:FREQuency:MODE
<Mode>
The command sets the instrument operating mode and therefore also the commands
used to set the output frequency.
Parameters:
<Mode>
CW | FIXed | SWEep
CW|FIXed
The instrument operates in fixed-frequency mode.
CW
and
FIXed
are synonyms.
The output frequency is set with
SWEep
The instrument operates in SWEep mode. The frequency is set
using the commands
or
*RST:
CW
SOURce Subsystem