Remote Control Commands
R&S
®
SMC100A
265
Operating Manual 1411.4060.32 ─ 04
Example:
LFO:FREQ:MODE SWE
sets the SWEep mode. The settings under
SOURce:LFOutput:FREQuency:STARt; STOP; MANual
become effective.
[:SOURce]:LFOutput:FREQuency:STARt
<Start>
This command sets the start frequency for the LF Sweep mode.
Parameters:
<Start>
float
Range:
0.1 Hz to 1 MHz
Increment: 0.1 Hz
*RST:
1 kHz
Example:
RST*
activates all presettings.
LFO:SWE:MODE AUTO
sets the
AUTO
sweep mode, i.e. each trigger triggers a complete
sweep.
TRIG0:SOUR SING
sets the
SINGle
trigger mode, i.e. the sweep is triggered by the
command
:LFOutput:SWEep:EXECute
or
*TRG
.
LFO:FREQ:STAR 100 kHz
sets the start frequency for the LF sweep to 100 kHz.
LFO:FREQ:STOP 200 kHz
sets the stop frequency of the LF sweep to 200 kHz.
LFO:FREQ:MODE SWE
sets the LF sweep mode.
LFO:SWE:EXEC
a one-off LF sweep from 100 kHz to 200 kHz is performed. The
linear step width is 1 kHz with a dwell time of 15 ms (preset values).
[:SOURce]:LFOutput:FREQuency:STOP
<Stop>
This command sets the stop frequency for the LF sweep.
Parameters:
<Stop>
float
Range:
0.1 Hz to 1 MHz
Increment: 0.1 Hz
*RST:
50 kHz
Example:
LFO:FREQ:STOP 200 kHz
sets the stop frequency for the LF sweep to 200 kHz.
LFO:FREQ:STAR 100 kHz
sets the start frequency for the LF sweep to 100 kHz.
SOURce Subsystem