Instrument Function
R&S
®
SMC100A
133
Operating Manual 1411.4060.32 ─ 04
Fill User Correction Data with Sensor Settings
The "Fill with Sensor" button of the "Edit User Correction Data" menu opens the associ-
ated menu.
The menu indicates the relevant generator settings.
Fill User Correction Data with Sensor
All settings are read-only, except the "Sensor". In case more than one sensor is connec-
ted to the instrument, the sensor can be selected in the "Sensor" field of menu.
The "Execute" button is only enabled if a sensor is detected and the user correction list
contains at least one frequency value.
SCPI command:
5.3.9 Sweep Mode
5.3.9.1
Overview
The R&S Signal Generator offers three different sweep types (frequency sweep, level
sweep and LF sweep) to be activated alternatively. Each type has 6 modes which differ
with respect to the sweep cycle mode (continuous, individual and step-by-step) and trig-
gering mode (automatic, internal and external).
Sweep modes can not be activated simultaneously, they deactivate each other.
Setting a sweep
A sweep is set in five basic steps which are shown below taking a frequency sweep as
an example.
RF Signal - RF Block