Remote Control Commands
R&S
®
SMC100A
247
Operating Manual 1411.4060.32 ─ 04
[:SOURce<hw>]:CORRection:CSET:DATA:POWer
<Power>
The command transfers the level data to the table selected
with
:CORRection:CSET:SELect
.
*RST
does not affect data lists. The numerical suffix at
SOURce
must not be used for this
command.
Parameters:
<Power>
Power#1[, Power#2, ...]
Range:
-40 dB to 6 dB
Default unit: dB
Example:
CORR:CSET '/var//Lists/ucor/ucor1'
selects the table
ucor1
.
CORR:CSET:DATA:POW 1dB, 0.8dB, 0.75dB,...
enters the level values in the table
ucor1
.
[:SOURce<hw>]:CORRection:CSET:DATA:POWer:POINts?
The command queries the number of level values in the selected table.
The numerical suffix at SOURce must not be used for this command.
Return values:
<Points>
float
Example:
CORR:CSET '/var//Lists/ucor/ucor1'
selects the table ucor1.
CORR:CSET:DATA:POW:POIN?
queries the number of level values in the table ucor1.
Response:
440
the table ucor1 contains 440 level values.
Usage:
Query only
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
The command fills the selected user correction list with the level values measured by the
power sensor for the given frequencies.
The power sensor used is selected by the suffix in key word
SENSe
of the command
header.
Suffix:
<ch>
.
1 .. 3
Suffix 1 denotes the sensor connected to the SENSOR connector,
suffix 2 the sensor connected first to one of the USB interfaces
and suffix 3 the sensor connected second to one of the USB inter-
faces.
SOURce Subsystem