Remote Control Commands
R&S
®
SMBV100B
647
User Manual 1178.4460.02 ─ 03
Suffix
Value range
Description
OUTPut<hw>
[1]
Optional suffix
............................................................................................647
.............................................................................................648
:OUTPut<hw>:AFIXed:RANGe:LOWer?
........................................................................... 648
:OUTPut<hw>:AFIXed:RANGe:UPPer?
............................................................................648
................................................................................... 649
:OUTPut<hw>:PROTection:TRIPped?
.............................................................................. 649
:OUTPut:ALL[:STATe]
<State>
Activates the RF output signal of the instrument.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
n.a. (factory preset: 0)
Example:
OUTPut:ALL:STATe 0
:OUTPut<hw>[:STATe]
<State>
Activates the RF output signal.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
OUTP ON
Activates the RF output.
Manual operation:
See
:OUTPut<hw>[:STATe]:PON
<Pon>
Defines the state of the RF output signal when the instrument is switched on.
Parameters:
<Pon>
OFF | UNCHanged
*RST:
n.a. (factory preset: UNCHanged)
Example:
OUTP:PON OFF
The RF output is deactivated when the instrument is switched
on.
Manual operation:
See
OUTPut Subsystem