Configuring the Internal Baseband Source
R&S
®
SMBV100B
127
User Manual 1178.4460.02 ─ 03
The dialog comprises the settings used to configure the power ramping, like the source
of the control signals, the form of the ramp function and the applied attenuation.
See also
"Power Ramping and Level Attenuation"
Power ramping is possible up to a symbol rate of 5 MHz. A higher symbol rate disables
the power ramping automatically and an error message is output.
Settings:
State
Enables/disables power ramping.
Remote command:
[:SOURce<hw>]:BB:DM:PRAMp[:STATe]
Source
Defines the source for the power ramp control signals.
"Internal"
The control signals as defined in the internal control list are used for
control purposes.
See also
Chapter 3.5.4.1, "How to Create and Assign a Control List"
Remote command:
[:SOURce<hw>]:BB:DM:PRAMp:SOURce
Ramp Function
Selects the ramp function that describes the shape of the rising and falling edges dur-
ing power ramp control, see
"Linear"
The transmitted power rises and falls linear fashion.
"Cosine"
The transmitted power rises and falls with a cosine-shaped edge,
resulting in a more favorable spectrum than the "Linear" setting.
Remote command:
[:SOURce<hw>]:BB:DM:PRAMp:SHAPe
Ramp Time
Enters the power ramping rise time and the fall time for a burst. The setting is
expressed in symbols. See also
Remote command:
[:SOURce<hw>]:BB:DM:PRAMp:TIME
Generating Custom Digital Modulated Signals