Remote Control Commands
R&S
®
SGU100A
167
User Manual 1176.7726.02 ─ 09
:MMEMory:STORe:STATe
<savrcl_state_nr>, <file_name>
Stores the current instrument setting in the specified file.
The instrument setting must first be stored in an internal memory with the same num-
ber using the common command
*SAV
.
Setting parameters:
<savrcl_state_nr>
Corresponds to the specific
<number>
defined with the
*SAV
command, e.g.
*SAV 4
.
<file_name>
String parameter to specify the file name with extension
*.savrcltxt
.
Example:
See
"Storing and Loading Current Settings"
Usage:
Event
11.11
OUTPut Subsystem
................................................................................... 168
....................................................................................168
................................................................................... 169
:OUTPut:AMODe
<AMode>
Sets the mode of the attenuator (
A
ttenuator
MOD
e) at the RF output.
Parameters:
<AMode>
AUTO | FIXed
AUTO
The attenuator is switched automatically. The level settings are
made in the full range.
FIXed
The level settings are made without switching the attenuator.
When this operating mode is switched on, the attenuator is fixed
to its current position and the resulting variation range is defined.
*RST:
AUTO
Manual operation:
See
:OUTPut[:STATe]
<State>
Activates/deactivates the RF output.
Parameters:
<State>
0 | 1 | OFF | ON