Remote Control Commands
R&S
®
SGU100A
169
User Manual 1176.7726.02 ─ 09
:OUTPut<hw>:PROTection:CLEar
Resets the protective circuit after it has been tripped. The state of the output is again
determined by
OUTPut:STATe
.
Example:
OUTP:PROT:CLE
resets the protective circuit for RF output.
Usage:
Event
11.12
SOURce Subsystem
[:SOURce]:SETTings:APPLy[:IMMediate]
......................................................................... 169
[:SOURce]:FREQuency[:CW|FIXed]
.................................................................................169
............................................................................................. 170
[:SOURce]:SETTings:APPLy[:IMMediate]
Applies the signal settings and outputs the signal.
See
Chapter 11.14, "SOURce:Local Oscillator Subsystem"
tion of the setting process.
Usage:
Event
Manual operation:
See
"Confirm External LO Settings"
[:SOURce]:FREQuency[:CW|FIXed]
<Cw>
Sets the RF frequency at the RF output connector of the instrument.
Parameters:
<Cw>
float
Range:
1E+6 to 40E+9
Increment: 1E-3
*RST:
1E+9
Example:
See
Chapter 11.1.2, "Generating an I/Q Modulated Signal"
Manual operation:
See
[:SOURce]:TRAits<ch>
Queries for a list of values for the corresponding trait. The following values values are
available for the channel number:
●
1: queries the upper edge frequencies for the frequency bands
●
2: queries the upconverter frequency multiplication factors for each band
●
3: queries if the bypass mode is active for each band
SOURce Subsystem