Preface
R&S
®
Scope Rider RTH
6
Getting Started 1326.1561.02 ─ 05
Input channels
●
R&S
RTH1002: 2 oscilloscope channels, 1 multimeter
●
R&S
RTH1004: 4 oscilloscope channels
Maximum input voltage
●
At BNC inputs: CAT
IV
300
V
●
With probe R&S RT-ZI10 or R&S RT-ZI11: CAT
IV
600
V, CAT
III
1000
V
●
Meter input: CAT
IV
600
V; CAT
III
1000
V
Voltage ratings: V RMS (50 to 60
Hz) for AC sine wave and V DC for DC applica-
tions.
Vertical system
●
Analog bandwidth (–3 dB) of oscilloscope channels:
– R&S
RTH1002 and R&S
RTH1004: ≥ 60 MHz
– R&S
RTH1002 with -B221 option and R&S
RTH1004 with -B241 option:
≥
100
MHz
– R&S
RTH1002 with -B222 option and R&S
RTH1004 with -B242 option:
≥
200
MHz
– R&S
RTH1002 with -B223 option and R&S
RTH1004 with -B243 option:
≥
350
MHz
– R&S
RTH1002 with -B224 option and R&S
RTH1004 with -B244 option:
≥
500
MHz
●
Rise time (calculated):
R&S
RTH1002 and R&S
RTH1004 with bandwidth of 60
MHz, 100
MHz,
200
MHz, 350
MHz, 500
MHz: <
5.8 ns, <
3.5 ns, <
1.75 ns, <
1 ns, <
700 ps
respectively
●
DC gain accuracy:
Offset and position set to zero after self-alignment
– Input sensitivity > 5 mV/div:
±
1.5 %
– Input sensitivity > 2 mV/div to 5 mV/div :
±
2 %
– input sensitivity 2 mV/div:
±
2,5 %
●
Input impedance: 1 M
Ω
±
1 % || 12 pF
±
2 pF (meas.)
●
Input coupling: DC, AC
●
Input sensitivity: 2 mV/div to 100 V/div
Main Specifications