Remote Control Commands
R&S
®
RTM3000
782
User Manual 1335.9090.02 ─ 08
POSitive
Rising edge (rise).
NEGative
Falling edge (fall).
EITHer
Rising as well as the falling edge (both).
*RST:
POSitive
16.14.2.2
Square Wave
PGENerator:PATTern:SQUarewave:POLarity
....................................................................782
PGENerator:PATTern:SQUarewave:DCYCle
.....................................................................782
..........................................................................................782
................................................................................... 783
PGENerator:PATTern:SQUarewave:POLarity
<Polarity>
Defines the polarity of the pattern generator square wave function.
Parameters:
<Polarity>
NORMal | INVerted
*RST:
NORM
PGENerator:PATTern:SQUarewave:DCYCle
<DutyCycle>
Sets the duty cycle of the square wave function.
Parameters:
<DutyCycle>
Numeric value
Range:
1.00E+00 to 9.900E+01
*RST:
5.000E+01
Default unit: %
Example:
PGEN:PATT:SQU:DCYC 20
Sets the duty cycle of the square wave function to 20%.
PGENerator:PATTern:PERiod
<PatternPeriod>
Defines the period of the pattern generator function.
Parameters:
<PatternPeriod>
Numeric value (Period = Pattern length * Bit time)
Range:
MIN 1 Sample * 20ns = 20ns to MAX 2048 Sam-
ples * 42s = 10416s (approx. 2.89h)
*RST:
2.000E-06
Default unit: s
Signal Generation (Option R&S
RTM-B6)
www.allice.de
Allice Messtechnik GmbH