Putting into Operation
R&S
®
RT
‑
Z1M
12
Manual 1801.3142.02 ─ 02
3.2
Identification of Adapter and Probe
When the R&S
RT
‑
Z1M is connected to the oscilloscope, the oscilloscope recognizes
the adapter and reads out the specific parameters from the data memory of the
adapter.
When a probe is connected to the adapter, the oscilloscope also reads the probe-spe-
cific parameters through the adapter.
For information on probe-specific settings and information, refer to the oscilloscope's
user manual.
3.3
Connecting the Adapter to the DUT
Observe the following step order when connecting the components of the measure-
ment setup. Also read and observe the instructions in
Chapter 1, "Safety Information"
,
on page 5.
1. Make sure that the oscilloscope is properly grounded.
2. Connect the adapter to the oscilloscope.
3. Connect the probe to the adapter.
4. Switch off the test circuit.
5. Connect the probe to the DUT. Ensure a stable connection between the DUT and
the probe.
6. Switch on the test circuit.
3.4
Disconnecting the Adapter from the DUT
Observe the following step order when disconnecting the measurement setup. Also
read and observe the instructions in
Chapter 1, "Safety Information"
,
on page 5.
1. Switch off the test circuit.
2. Disconnect the probe from the DUT.
3. Disconnect the probe from the adapter.
4. Disconnect the adapter from the oscilloscope.
3.5
Using the Adapter
All settings of the R&S
RT
‑
Z1M are adjusted on the oscilloscope.
Using the Adapter