![Rohde & Schwarz R&S FSP Series Operating Manual Download Page 338](http://html.mh-extra.com/html/rohde-and-schwarz/rands-fsp-series/rands-fsp-series_operating-manual_1477864338.webp)
R&S FSP
Instrument Functions
LO/IF ports for external mixers - Option R&S FSP-B21
4.266
Operating Manual 1164.4556.12 - 06
Two-port mixer:
➢
Connect the 'LO OUT / IF IN' output of the R&S FSP to the LO/IF port of the
external mixer. The nominal LO level is 15.5 dBm.
Because of the diplexer contained in the R&S FSP the IF signal can be tapped
from the line which is used to feed the LO signal to the mixer.
➢
Feed the signal to be measured to the RF input of the external mixer.
4.13.2
Manual Operation
The
FREQ
hardkey opens the menu for selecting the frequency and the frequency
range (details see
“Frequency and Span Selection – FREQ Key” on page 4.9
extended by the following functions for external mixing:
FREQ
S P E C T R U M A N A L Y Z E R 9 k H z . .. 3 G H z F S P
.
1 0 9 3 .4 4 9 5 . 0 3
.
M A D E I N GE R MA N Y
P R E S E T
M A D E I N GE R MA N Y
L O O U T / IF IN
E x te rn a l
M ix e r
R F IN P U T
R F
L O / IF
!
!
!
!
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...