R&S FSP
Instrument Functions
LO/IF ports for external mixers - Option R&S FSP-B21
4.270
Operating Manual 1164.4556.12 - 06
After selection of a file the parameters
BIAS
for the
SELECT BAND
table are taken
from that file. The selected mixer type must correspond to the file entry
PORTS
. It is
not possible to modify these parameters during operation.
Only that part of the table is taken into account that contains reference values for fre-
quencies that are covered by the selected order of harmonic. (See section
sion Loss Tables” on page 4.271
Remote command:
MIX:LOSS:TABL mix_1_4
CONV LOSS
TABL
The
CONV LOSS TABLE
softkey opens a submenu for selecting and editing of con-
version loss tables. (See section
“Conversion Loss Tables” on page 4.271
)
Remote command:
--
ACCEPT BIAS
The
ACCEPT BIAS
softkey stores the bias setting selected in the table
SELECT
BAND
in the file specified in the table.
The user is prompted to confirm the storage. The new setting will be stored only
when confirmed with
YES
.
This softkey can only be used if
CONV LOSS
TABLE
(conversion loss taken into
account by means of frequency-dependent values) is active.
Remote command:
--
LO LEVEL
The softkey
LO LEVEL
changes the LO level of the external mixer LO port from 13.0
dBm to 17.0 dBm in 0.1 dB steps. Default value is 15.5 dB.
Remote command:
[SENSe1:]MIXer:LOPower 16.0dBm
SIGNAL ID /
AUTO ID /
AUTO ID
THRESHOLD
The
SIGNAL ID
,
AUTO ID
and
AUTO ID THRESHOLD
softkeys are used for signal
identification.
Remote command:
MIX:SIGN ON
See section
“Signal Identification” on page 4.277
.
Bias Current
Single-diode mixers generally require a DC voltage which is applied via the LO line.
This DC voltage is to be tuned to the minimum conversion loss versus frequency.
Such a DC voltage can be set via the
BIAS
function using the D/A converter of the
R&S FSP. The value to be entered is not the voltage but the short-circuit current.
<NONE>
Band_A_4
Mixer_4
SELECT
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...