Remote Commands
R&S
®
Spectrum Rider FPH
323
User Manual 1321.1011.02 ─ 06.00
Example:
INIT:CONT OFF
Turns on single sweep mode.
CALC:MARK2 ON
CALC:MARK2:FUNC:NOIS ON
Turns on marker 2 and assigns the noise measurement to that
marker.
INIT;*WAI
CALC:MARK2:NOIS:RES?
Performs the measurement and queries the noise marker
results.
Usage:
Query only
CALCulate<1...2>:MARKer<1...6>:FUNCtion:REFerence
This command matches the reference level to the power level of a marker.
If you use a delta marker, the R&S Spectrum Rider turns it into a normal marker.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Example:
CALC:MARK1:FUNC:REF
Matches the reference level to the power level of marker 1.
Usage:
Event
13.6.2.7
Using Limit Lines
The following commands define limit lines and perform the corresponding limit checks.
List of commands
●
CALCulate<1...2>:LIMit<1...2>:BEEP[:STATe]
●
CALCulate<1...2>:LIMit<1...2>:COMMent?
●
CALCulate<1...2>:LIMit<1...2>:DEFine
●
CALCulate<1...2>:LIMit<1...2>:DELete
●
●
●
CALCulate<1...2>:LIMit<1...2>:FAIL?
●
CALCulate<1...2>:LIMit<1...2>:STATe
●
CALCulate<1...2>:LIMit<1...2>:UNIT:X?
●
CALCulate<1...2>:LIMit<1...2>:UNIT[:Y]?
●
CALCulate<1...2>:LIMit<1...2>:LOWer:SELect
●
CALCulate<1...2>:LIMit<1...2>:UPPer:SELect
●
CALCulate<1...2>:LIMit<1...2>:LOWer:THReshold
Remote Control - Commands