Remote Commands
R&S
®
Spectrum Rider FPH
318
User Manual 1321.1011.02 ─ 06.00
CALCulate<1...2>:MARKer<1...6>:COUNt:FREQuency?
This command performs a frequency measurement at the marker position and returns
the result.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end to make sure that the R&S Spectrum Rideractually reaches the frequency
you want to measure. This is only possible in single sweep mode.
Before you can use the command, you have to turn on the frequency counter with
CALCulate<1...2>:MARKer<1...6>:COUNt[:STATe]
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Example:
See
CALCulate<1...2>:MARKer<1...6>:COUNt[:STATe]
Usage:
Query only
CALCulate<1...2>:MARKer<1...6>:COUNt[:STATe]
<State>
This command turns the frequency counter at the marker position on and off.
You can read out the result with
CALCulate<1...2>:MARKer<1...6>:COUNt:
Frequency counting is possible only for one marker at a time. If it is activated for
another marker, it is automatically deactivated for the previous marker.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end to make sure that the R&S Spectrum Rider actually reaches the frequency
you want to measure. This is only possible in single sweep mode.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
INIT:CONT OFF
CALC:MARK ON
Turns on single sweep mode and marker 1.
CALC:MARK:COUN ON
Turns on the frequency counter for marker 1.
INIT;*WAI
CALC:MARK:COUN:FREQ?
Performs a measurement and queries the results of the fre-
quency counter.
CALCulate<1...2>:MARKer<1...6>:FREQuency:MODE
<Mode>
This command selects the marker frequency display mode.
Remote Control - Commands