The accuracy of measurement results can be impaired because the oven-control-
led crystal oscillator has a too low temperature. After switching-on the instrument,
the OCXO requires a warm-up phase to reach its operating temperature.
●
101 ("Firmware Error")
:
Indicates a firmware or software error. If you encounter this error for the first time,
restart the instrument.
If the error occurs again, consider the following hints:
–
Firmware errors can often be repaired by restoring the factory default settings.
–
If a software package has not been properly installed, this failure is often indi-
cated in the "Setup" dialog, section "SW/HW-Equipment > Installed Software".
–
Check for software updates correcting the error. Updates are provided in the
customer web on GLORIS (registration required):
If you get firmware errors even with the properly installed latest software version,
send a problem report including log files to Rohde & Schwarz.
●
102 ("Unidentified Error")
:
Indicates an error not covered by other reliability values. For troubleshooting, follow
the steps described for "101 (firmware error)".
●
103 ("Parameter Error")
:
Indicates that the measurement could not be performed due to internal conflicting
parameter settings.
A good approach to localize the conflicting settings is to start with a reset or preset
or even restore the factory default settings. Then reconfigure the measurement
step by step and check when the error occurs for the first time.
If you need assistance to localize the conflicting parameter settings, contact Rohde
& Schwarz (see
http://www.service.rohde-schwarz.com
●
104 ("Not Functional")
:
The application could not be started with the configured parameter set.
2.5.2
Connection Control and States
The commands in the following sections control the connection to the UE and retrieve
connection-related information.
SOURce:NIOT:SIGN<i>:CELL:STATe
............................................................................... 111
SOURce:NIOT:SIGN<i>:CELL:STATe:ALL?
.......................................................................112
................................................................................... 112
CALL:NIOT:SIGN<i>:PSWitched:ACTion
.......................................................................... 113
FETCh:NIOT:SIGN<i>:PSWitched:STATe?
........................................................................113
SOURce:NIOT:SIGN<i>:CELL:STATe
<Control>
Turns the generator (the cell) on or off.