Functions
R&S
®
OSP
198
User Manual 1178.7100.02 ─ 05
Switch-On Action
Specifies, if a path is switched after booting the R&S
OSP. The following options are
available:
"None"
The R&S
OSP switches no path at instrument startup.
"Switch Path"
Upon booting, the R&S
OSP switches the path specified in the
field.
If a switch unit is in
, the setting in the primary switch
unit determines, if the secondary device executes a "Switch-On
Action" at Power Up.
Path ← Switch-On Action
Specifies the path to be switched at startup.
Only available, if the "Switch-On Action" is "Switch Path".
From the path list that opens upon clicking the "Path" field, select one of the existing
paths.
Note that the dialog also offers non-switchable paths. In the
can check, whether a path is switchable or not.
If you select a non-switchable path, the R&S
OSP does not perform any switching
action at startup.
Remote command:
Path Info
Enables or disables in the main page the line "Last Switched Path", showing one of the
following pieces of information:
●
Path_XY
(the name of the previously switched path)
●
<Reset State>
, if the switch unit is in reset state
See also:
–
–
–
●
<Individual Settings>
, if any relays or channels in the switch unit were
switched via the web interface or by a remote control command. Such a modifica-
tion can result in a previously switched path losing its integrity. Therefore, even if
actual path integrity is not checked by the firmware, it considers the modules to be
in any individual switching state.
Remote command:
CONFigure:MAINinfo:PATH
ROUTe:PATH:LAST?
Customer Text
Specifies the string displayed as "Customer Text" in the "Main" page, see
Remote command:
Configuration