Remote control commands
R&S
®
NRPM
155
Mannual 1425.8663.02 ─ 08
Example:
[SENSe1:][POWer:][AVG:]BUFFer:CLEar
Usage:
Event
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:COUNt?
Available in continuous average mode.
Queries the number of results that are currently saved in the result buffer.
Example:
[SENSe1:][POWer:][AVG:]BUFFer:COUNt?
Usage:
Query only
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:DATA?
Queries the data of the continuous average result buffer and returns them even if
the buffer is not full.
FETCh<Channel>[:SCALar][:POWer][:AVG]?
only if the buffer is full.
Example:
[SENSe1:][POWer:][AVG:]BUFFer:DATA?
Usage:
Query only
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:INFO?
[<ITEM>]
Queries the data of the continuous average result buffer.
Query parameters:
<ITEM>
Example:
[SENSe1:][POWer:][AVG:]BUFFer:INFO?
Usage:
Query only
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:SIZE
<count>
Sets the size of the result buffer in continuous average mode.
You can enable the buffer using
Configuring the measurement modes