Remote control commands
R&S
®
NRPM
150
Mannual 1425.8663.02 ─ 08
FETCh<Channel>[:SCALar][:POWer][:AVG]?
Queries the measurement results of a particular channel of the sensor module.
Suffix:
<Channel>
.
1...3
Usage:
Query only
FETCh<Sensor>:ALL[:SCALar][:POWer][:AVG]?
Queries the measurement results of all channels of a sensor module.
The sensor module returns the results in a string, separated by commas.
Usage:
Query only
CALCulate:FEED
<mode>
When you query measurement data using
, the sensor module returns data of the measurand that was
configured before, but it can also return data of different measurands. By default,
the reading is the average power.
To determine the measurand for the
FETCh<Channel>[:SCALar][:POWer][:
command, use the
CALCulate:FEED
before initiating the measurement.
Depending on the measurement mode, the following settings are possible:
SENS:FUNC
Possible CALC:FEED
Meaning
"POWer:AVG"
"POWer:AVERage"
Average value
"POWer:PEAK"
Peak value
"POWer:RANDom"
Randomly selected value from
the measurement interval
"XTIMe:POWer"
"POWer:TRACe"
Measurement sequence
"POWer:PEAK:TRACe"
Peak value of the samples per
trace point
"POWer:RANDom:TRACe"
Randomly selected value of
the samples per trace point
Parameters:
<mode>
*RST:
"POWer:AVERage"
Selecting a measurement mode and retrieving results