Data Management
R&S
®
FSW
560
User Manual 1173.9411.02 ─ 19
File contents
Description
Sweep Count;20;
Number of sweeps set
Data section
Trace 1:;;
Selected trace
x-Unit;Hz;
Unit of x values: Hz with span > 0; s with span = 0; dBm/dB
with statistics measurements
y-Unit;dBm;
Unit of y values: dB*/V/A/W depending on the selected unit
with y-axis LOG or % with y-axis LIN
Values; 1001;
Number of measurement points
Frames;2;
Number of exported frames
Frame;0;
Most recent frame number
Timestamp;17.Mar 11;11:27:05.990
Timestamp of this frame
10000;-10.3;-15.7
10130;-11.5;-16.9
10360;-12.0;-17.4
...;...;
Measured values, identical to spectrum data:
<x value>, <y1>, <y2>; <y2> being available only with detector
AUTOPEAK and containing in this case the smallest of the two
measured values for a measurement point.
Frame;-1;
Next frame
Timestamp;17.Mar 11;11:27:05.342
Timestamp of this frame
...
8.5 Creating Screenshots of Current Measurement
Results and Settings
In order to document the graphical results and the most important settings for the cur-
rently performed measurement, you can create a hardcopy or screenshot of the current
display. Screenshots can either be printed or stored to a file.
...............................................................................560
How to Store or Print Screenshots of the Display
.................................................564
8.5.1 Print and Screenshot Settings
The settings for saving and printing screenshots are configured via the "Print" menu
which is displayed when you select the "Print" icon in the toolbar.
For step-by-step instructions see
chapter 8.5.2, "How to Store or Print Screenshots of
Remote commands for these settings are described in
chapter 11.9.4, "Storing or Print-
Creating Screenshots of Current Measurement Results and Settings