Common Measurement Settings
R&S
®
FSW
328
User Manual 1173.9411.02 ─ 19
For details on using a power sensor as an external trigger, see
Remote command:
TRIG:SOUR PSE
, see
External Trigger Level ← Using the power sensor as an external trigger
Defines the trigger level for the power sensor trigger.
For details on supported trigger levels, see the data sheet.
Remote command:
Hysteresis ← Using the power sensor as an external trigger
Defines the distance in dB to the trigger level that the trigger source must exceed
before a trigger event occurs. Setting a hysteresis avoids unwanted trigger events
caused by noise oscillation around the trigger level.
Remote command:
Trigger Holdoff ← Using the power sensor as an external trigger
Defines the minimum time (in seconds) that must pass between two trigger events.
Trigger events that occur during the holdoff time are ignored.
Remote command:
Drop-Out Time ← Using the power sensor as an external trigger
Defines the time the input signal must stay below the trigger level before triggering
again.
Slope ← Using the power sensor as an external trigger
Defines whether triggering occurs when the signal rises to the trigger level or falls
down to it.
Remote command:
6.2.3.3
How to Work With a Power Sensor
The following step-by-step instructions demonstrate how to set up a power sensor. For
details on individual functions and settings see
chapter 6.2.3.2, "Power Sensor Set-
The remote commands required to perform these tasks are described in
ter 11.7.6.5, "Working with Power Sensors"
Data Input and Output