Common Measurement Settings
R&S
®
FSW
323
User Manual 1173.9411.02 ─ 19
Using a Power Sensor as an External Power Trigger
Power sensors can be used to trigger a measurement at a specified power level, e.g.
from a signal generator.
Currently, only the following power sensors are supported as power triggers:
●
R&S NRP-Z81
●
R&S NRP-Z85
●
R&S NRP-Z86
With the R&S
FSW, the power sensors can be connected to the "Power Sensor" inter-
face directly, and no further cables are required. They can then be configured as an
external power sensor trigger.
Fig. 6-2: Connecting a power sensor using the POWER SENSOR interface
The R&S
FSW receives an external trigger signal when the defined trigger level is
measured by the power sensor. Power measurement results are provided as usual.