Analysis
R&S
®
FSW-K40
58
User Manual 1173.9286.02 ─ 15
Remote command:
Selecting a trace:
CALCulate<n>:EVALuation:USER<range>:TRACe
Define start frequency of user range:
CALCulate<n>:EVALuation:USER<range>:STARt
Define stop frequency of user range:
CALCulate<n>:EVALuation:USER<range>:STOP
6.2.2 Configuring Spot Noise Measurements
On User Defined Offsets / Offset Frequency
................................................................ 58
On All Decade Edges
Turns the calculation of spot noise on all 10
x
offset frequencies on and off.
Remote command:
Turn on and off spot noise calculation on 10
x
offset frequencies:
CALCulate<n>:SNOise:DECades[:STATe]
Querying spot noise results on 10
x
offset frequencies:
CALCulate<n>:SNOise:DECades:X?
CALCulate<n>:SNOise:DECades:Y?
On User Defined Offsets / Offset Frequency
Turns custom spot noise frequencies on and off.
If on, the "Offset Frequency" input fields become available. You can measure the spot
noise for up to five custom offset frequencies. If active, the application adds those
spots to the spot noise table.
Remote command:
Turning spot noise marker on and off:
Positioning spot noise markers:
Querying custom spot noise results:
6.3 Using Limit Lines
Access
: "Overview" > "Limit Analysis"
The "Limits" tab of the "Results" dialog box contains all funtions necessary to set up
and configure limit lines.
Using Limit Lines