Remote Commands
R&S
®
FPL1000
770
User Manual 1178.3370.02 ─ 03
CALCulate<n>:LIMit<k>:UPPer[:DATA]
............................................................................776
CALCulate<n>:LIMit<k>:UPPer:MARGin
.......................................................................... 776
CALCulate<n>:LIMit<k>:UPPer:MODE
.............................................................................776
CALCulate<n>:LIMit<k>:UPPer:OFFSet
........................................................................... 777
CALCulate<n>:LIMit<k>:UPPer:SHIFt
.............................................................................. 777
CALCulate<n>:LIMit<k>:UPPer:SPACing
......................................................................... 778
CALCulate<n>:LIMit<k>:UPPer:STATe
............................................................................ 778
CALCulate<n>:LIMit<k>:UPPer:THReshold
...................................................................... 778
CALCulate<n>:LIMit<k>:COMMent
<Comment>
This command defines a comment for a limit line.
Suffix:
<n>
.
irrelevant
<k>
Parameters:
<Comment>
String containing the description of the limit line.
Manual operation:
See
CALCulate<n>:LIMit<k>:CONTrol[:DATA]
<LimitLinePoints>
This command defines the horizontal definition points of a limit line.
Suffix:
<n>
.
irrelevant
<k>
Parameters:
<LimitLinePoints>
Variable number of x-axis values.
Note that the number of horizontal values has to be the same as
the number of vertical values set with
FPL1000 either adds missing
values or ignores surplus values.
The unit is Hz or s.
*RST:
-
Manual operation:
See
CALCulate<n>:LIMit<k>:CONTrol:DOMain
<Type>
This command selects the domain of the limit line.
Suffix:
<n>
.
irrelevant
<k>
Setting Basic Spectrum RF Measurement Parameters