
Remote Commands
R&S
®
FPL1000
777
User Manual 1178.3370.02 ─ 03
Suffix:
<n>
.
<k>
Parameters:
<Mode>
ABSolute
Limit line is defined by absolute physical values.
The unit is variable.
RELative
Limit line is defined by relative values related to the reference
level (dB).
*RST:
ABSolute
Manual operation:
See
CALCulate<n>:LIMit<k>:UPPer:OFFSet
<Offset>
This command defines an offset for a complete upper limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Suffix:
<n>
.
irrelevant
<k>
Parameters:
<Offset>
Numeric value.
*RST:
0
Default unit: dB
Manual operation:
See
CALCulate<n>:LIMit<k>:UPPer:SHIFt
<Distance>
This command moves a complete upper limit line vertically.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
.
irrelevant
<k>
Parameters:
<Distance>
Defines the distance that the limit line moves.
The unit depends on
Usage:
Event
Manual operation:
See
Setting Basic Spectrum RF Measurement Parameters