Remote Commands
R&S
®
FPL1000
772
User Manual 1178.3370.02 ─ 03
CALCulate<n>:LIMit<k>:CONTrol:SHIFt
<Distance>
This command moves a complete limit line horizontally.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
.
irrelevant
<k>
Parameters:
<Distance>
Numeric value.
The unit depends on the scale of the x-axis.
Manual operation:
See
CALCulate<n>:LIMit<k>:CONTrol:SPACing
<InterpolMode>
This command selects linear or logarithmic interpolation for the calculation of limit lines
from one horizontal point to the next.
Suffix:
<n>
.
<k>
Parameters:
<InterpolMode>
LINear | LOGarithmic
*RST:
LIN
Example:
CALC:LIM:CONT:SPAC LIN
Manual operation:
See
CALCulate<n>:LIMit<k>:LOWer[:DATA]
<LimitLinePoints>
This command defines the vertical definition points of a lower limit line.
Suffix:
<n>
.
irrelevant
<k>
Parameters:
<LimitLinePoints>
Variable number of level values.
Note that the number of vertical values has to be the same as
the number of horizontal values set with
FPL1000 either
adds missing values or ignores surplus values.
The unit depends on
on page 775.
*RST:
Limit line state is OFF
Setting Basic Spectrum RF Measurement Parameters