Remote Commands
R&S
®
FPL1000
690
User Manual 1178.3370.02 ─ 03
Manual operation:
See
" Trace 1 / Trace 2 / Trace 3 / Trace 4 / Trace 5 / Trace 6 "
on page 467
See
" Trace 1 / Trace 2 / Trace 3 / Trace 4 (Softkeys)"
DISPlay[:WINDow<n>]:TRACe<t>:SMOothing:APERture
<Percentage>
This command defines the degree (aperture) of the trace smoothing, if
WINDow<n>]:TRACe<t>:SMOothing[:STATe]
TRUE
.
Suffix:
<n>
.
<t>
Parameters:
<Percentage>
Range:
1 to 50
*RST:
2
Default unit: PCT
Example:
DISP3:TRAC2:SMO:APER 5
Defines an aperture of 5% for trace 2 in window 3
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:SMOothing[:STATe]
<State>
This command turns trace smoothing for a particular trace on and off.
If enabled, the trace is smoothed by the value specified using
WINDow<n>]:TRACe<t>:SMOothing:APERture
For more information see
Suffix:
<n>
.
<t>
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
DISP3:TRAC2:SMO ON
Turns on trace smoothing for trace 2 in window 3
Manual operation:
See
[SENSe:]AVERage<n>:COUNt
<AverageCount>
This command defines the number of sweeps that the application uses to average
traces.
Setting Basic Spectrum RF Measurement Parameters