
Remote Control
R&S
®
FPC
300
User Manual 1178.4130.02 ─ 07
Example:
//Select CISPT bandwidth
BAND:CISP 9KHZ
Manual operation:
See
[SENSe:]BANDwidth[:RESolution]:CISPr:AUTO
<State>
This command turns automatic selection of the CISPR measurement bandwidth on
and off.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
OFF
Example:
//Select CISPR bandwidth manually
BAND:CISP:AUTO OFF
BAND:CISP 120KHZ
Manual operation:
See
19.13.5
Sweep Control
Commands to control the measurement described elsewhere.
●
Measurement control:
●
Measurement type (single / continuous):
●
Measurement time:
19.13.6
Traces
Commands to configure traces described elsewhere.
●
Trace mode:
DISPlay<n>[:WINDow]:TRACe<t>:MODE
●
Memory trace:
DISPlay<n>[:WINDow]:TRACe<t>:MEMory[:STATe]
●
Data format of trace export:
[SENSe:]DETector<t>[:FUNCtion]
.................................................................................... 300
...................................................................................301
[SENSe:]DETector<t>[:FUNCtion]
<Detector>
This command selects the detector.
Remote Commands of the Receiver Application