Remote Control
R&S
®
FPC
269
User Manual 1178.4130.02 ─ 07
Manual operation:
See
See
See
See
FORMat[:DATA]
<DataFormat>
This command selects the data format that is used for transmission of trace data from
the R&S
FPC to the controlling computer.
Note that the command has no effect for data that you send to the R&S
FPC. The
R&S
FPC automatically recognizes the data it receives, regardless of the format.
Parameters:
<DataFormat>
ASCii
Returns the data in ASCII format, separated by commas.
REAL,32
Returns the data as 32-bit IEEE 754 floating point numbers in
the "definite length block format".
*RST:
ASCii
In REAL,32 format, a string of return values would look like:
#42524<value 1><value 2>...<value n>
with:
•
#4: Number of digits of the following number of data bytes (= 4
in this example).
•
2524: Number of following data bytes (2524, corresponds to
the 631 sweep points of the R&S
FPC.
•
<value>: 4-byte floating point value.
Example:
//Select the ASCii data format
FORM ASC
[SENSe:]DETector<t>[:FUNCtion]
<Detector>
This command selects the detector.
Suffix:
<t>
.
Parameters:
<Detector>
POSitive | NEGative | SAMPle | RMS | AVERage | APEak
For more information about the detectors, see
*RST:
APEak
Example:
//Select detector
DET POS
Manual operation:
See
Remote Commands of the Spectrum Application