![Rohde & Schwarz FPC Series User Manual Download Page 244](http://html.mh-extra.com/html/rohde-and-schwarz/fpc-series/fpc-series_user-manual_1477824244.webp)
Remote Control
R&S
®
FPC
241
User Manual 1178.4130.02 ─ 07
Parameters:
<Percentage>
<numeric value>
Range:
10 to 99.9%
*RST:
99%
Example:
//Define power percentage
CALC:MARK:FUNC:OBW:BAND:PCT 95
Manual operation:
See
19.12.1.5
TDMA Power
The following remote commands are exclusive to the TDMA power measurement.
Commands to configure TDMA power measurements described elsewhere.
●
Sweep time:
CALCulate<n>:MARKer<m>:FUNCtion:TDMA:BURSt
....................................................... 241
CALCulate<n>:MARKer<m>:FUNCtion:TDMA:BURSt
<BurstLength>
This command defines the burst length of the TDMA signal.
Prerequisites for this command
●
Turn on TDMA power measurement (
CALCulate<n>:MARKer<m>:FUNCtion:
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<BurstLength>
<numeric value>
The value range depends on the current sweep time.
*RST:
470 us
Default unit: s
Example:
//Define the burst length
CALC:MARK:FUNC:TDMA:BURS 80 US
Manual operation:
See
19.12.1.6
Harmonic Distortion
The following remote commands are exclusive to the harmonic distortion measure-
ment.
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:DISTortion?
........................................ 242
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:LIST?
.................................................242
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:NHARmonics
......................................243
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:PRESet
..............................................243
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics[:STATe]
.............................................. 244
Remote Commands of the Spectrum Application