Remote Control
R&S
®
FPC
242
User Manual 1178.4130.02 ─ 07
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:DISTortion?
<ResultType>
This command queries the total harmonic distortion of the signal.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end between activating the delta marker and reading out the result. This is only
possible in single sweep mode.
Prerequisites for this command
●
Turn on the harmonic distortion measurement (
Suffix:
<n>, <m>
.
irrelevant
Query parameters:
<ResultType>
TOTal
Return values:
<THD>
Pair of values, one showing the THD in %, one in dB.
Example:
See
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:
Usage:
Query only
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:LIST?
This command queries the position of the harmonics.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end between activating the delta marker and reading out the result. This is only
possible in single sweep mode.
Prerequisites for this command
●
Turn on the harmonic distortion measurement (
Suffix:
<n>, <m>
.
irrelevant
Return values:
<Position>
<numeric value>
Returns one value for every harmonic.
The first value is the absolute power of the first harmonic. The
level unit depends on the unit you have set with the
command. The other values are power levels relative to
the first harmonic. The unit for these is dB.
The total number of return values depends on the number of
harmonics you have set with
Remote Commands of the Spectrum Application