MMEMory:SELect:CHANnel[:ITEM]:TRANsducer:ALL
<State>
MMEMory:SELect[:ITEM]:TRANsducer:ALL
<arg0>
This command includes or excludes transducer factors when storing or loading a con-
figuration file.
Parameters:
<arg0>
ON | OFF | 1 | 0
*RST:
0
Example:
MMEM:SEL:TRAN:ALL ON
Manual operation:
See
15.8.3
Instrument Setting Management
....................................................................................... 591
..............................................................................592
........................................................................................ 592
................................................................................. 593
MMEMory:CLEar:ALL
This command deletes all <instrument> configuration files in the current directory.
You can select the directory with
Example:
MMEM:CLE:ALL
MMEMory:CLEar:STATe
<arg0>, <arg1>
This command deletes an instrument configuration file.
Parameters:
<arg0>
<arg1>
String containing the path and name of the file to delete.
The string may or may not contain the file's extension.
Example:
MMEM:CLE:STAT 1,'TEST'
MMEMory:LOAD:AUTO
<arg0>, <arg1>
This command restores an <instrument> configuration and defines that configuration
as the default state.
The default state is restored after a preset (
) or after you turn on the R&S
ESW.