Remote Commands in the Receiver Application
R&S
®
ESW
591
User Manual 1177.6298.02 ─ 08
Example:
MMEM:SEL:ALL
//Save all items (User Settings, All Traces, All Limit Lines) from
the R&S
ESW.
MMEM:LOAD:STAT 1,'C:\R_S\Instr\User\TEST01'
//Reloads all items
In the "Recall" dialog, select only "User Settings" and "All Limit
Lines".
MMEM:LOAD:STAT 1,'C:\R_S\Instr\User\TEST01'
//Reloads user settings and all limit lines.
*RST
//Reset <instrument>.
MMEM:LOAD:STAT 1,'C:\R_S\Instr\User\TEST01'
//Selected items are retained. Reloads user settings and all limit
lines.
Restart the <instrument>.
(Switch the [ON/OFF] key off and on).
MMEM:LOAD:STAT 1,'C:\R_S\Instr\User\TEST01'
// Selected items are set to default. Reloads only the user set-
tings.
Manual operation:
See
See
"Recall in New Channel / Recall in Current Channel"
MMEMory:STORe<1|2>:STATe
<arg0>, <arg1>
This command saves the current <instrument> configuration in a *.dfl file.
Secure User Mode
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details, see
Chapter 4.7, "Protecting Data Using the Secure User Mode"
Suffix:
<1|2>
.
irrelevant
Parameters:
<arg0>
<arg1>
String containing the path and name of the target file.
The file extension is .dfl.
Example:
MMEM:STOR:STAT 1,'Save'
Saves the current <instrument> settings in the file
Save.dfl
.
Manual operation:
See
Data Management