Service
R&S
®
ESH3-Z6
18
Manual 1178.9932.02 ─ 01
VNA
Port 1
Port 2
10 dB
10 dB
50
Short
R&S ESH3-Z6
X2.1
X2.2
X1.1
X1.2
X3
Figure 5-5: Test setup 1 for voltage division factor X2 - X3
Performing the test, part 1
The test is a S21 measurement on the vector network analyzer.
► Normalize Port 1 - Port 2 of the vector network analyzer.
Preparing the EUT, part 2
1.
DANGER!
Do not connect the R&S
ESH3-Z6 to mains voltage for the test. Con-
necting the R&S
ESH3-Z6 to the mains voltage poses hazardous or even life-
threatening risks to the operator.
Arrange the test set-up as shown in
.
2. Connect the Port 2 through the BNC 10
dB attenuator to the X3 connector.
3. Connect the BNC 50
Ω
termination to the BNC T – Branch input.
VNA
Port 1
Port 2
10 dB
50
Short
R&S ESH3-Z6
X2.1
X2.2
X1.1
X1.2
X3
10
dB
Figure 5-6: Test setup 1 for voltage division factor X2 - X3
Performing the test, part 2
► Read out the voltage division factor of conductor in frequency range 100
kHz to
200
MHz.
Target values:
● 100 kHz to 150 kHz: S21
≤
1.5
dB to
≤
1
dB (linearly decreasing with logarithm
of frequency)
● 150 kHz to 100 MHz: S21
≤
1 dB
● 100 MHz to 200 MHz: S21
≤
1
dB to
≤
2
dB (linearly increasing with logarithm
of frequency)
Perfomance Test