Command Reference
R&S
®
CMW500
199
User Manual 1173.9463.02 ─ 02
Example:
See
Usage:
Event
Firmware/Software:
V2.1.10
MMEMory:LOAD:MACRo
'<MacroLabel>', '<MacroFile>'
Loads a macro file, assigning an identifier <MacroLabel> to the macro file contents. Mac-
ros are deleted when the remote connection is closed. Macro files can be created e.g.
using the
command.
Note:
Avoid using macro labels which are identical with any of the remote control com-
mands supported by the R&S CMW500. In contrast to SCPI specifications, remote com-
mands have priority over macros.
Parameters:
'<MacroLabel>'
String parameter, used to execute the loaded macro and to refer-
ence it in other macro commands (e.g.
*DMC
,
*GMC?
,
*RMC
).
'<MacroFile>'
String parameter specifying path and filename of the source file
(on the instrument)
Example:
See
Usage:
Event
Firmware/Software:
V2.1.10
MMEMory:STORe:MACRo
'<MacroName>', '<MacroFile>'
Stores the contents of a macro identified by its <MacroName> to a file.
Parameters:
'<MacroName>'
String parameter, used to execute the loaded macro and to refer-
ence it in other macro commands (e.g.
*DMC
,
*GMC?
,
*RMC
).
'<MacroFile>'
String parameter specifying path and filename of the destination
file (on the instrument)
Example:
See
Usage:
Event
SCPI confirmed
Firmware/Software:
V2.1.10
7.3.2 Buffer Commands
The following remote control commands are related to response buffers. The command
syntax follows the general R&S CMW500 syntax rules; the commands are not SCPI
confirmed.
...............................................................................................200
Instrument-Control Commands