GUI Reference
R&S
®
CMW-KM750/-KM751
43
User Manual 1173.9670.02 ─ 05
The assembly level is ignored as long as the two-shot mode is disabled.
Remote command:
BER > Data Pattern
Selects the expected data pattern for looped-back RX test signals. Set this parameter
according to the data pattern used for generating the downlink signal (the ARB file).
The data pattern consists of zeros (ALL 0), ones (ALL 1), alternating zero one (01), alter-
nating one zero (10) or pseudo-random bit sequences of variable length (PN9, PN15).
Remote command:
BER > TrCH Mode
Selects the expected transport channel configuration for looped-back RX test signals.
Set this parameter according to the downlink signal (the ARB file).
"RMC"
The downlink signal contains both the Dedicated Traffic Channel
(DTCH) and the Dedicated Control Channel (DCCH).
"USER"
The downlink signal contains the DTCH, but no DCCH.
Remote command:
Assign Views
Selects the view types to be displayed in the overview. The R&S
CMW does not evaluate
the results for disabled views. Therefore, limiting the number of assigned views can speed
up the measurement.
Remote command:
List Mode Enable
Shows whether the list mode is enabled and disables an enabled list mode. Enabling the
list mode is only possible via the remote control command below. The list mode is essen-
tially a remote control feature; for an introduction see
chapter 3.3, "Multi Evaluation List
Parameters and Settings