Command Reference
R&S
®
CMW-B400B/-B405A
89
User Manual 1176.8116.02 ─ 04
<Level>
Level of the tone.
For method SEP, the level can be set and queried.
For method TLEV, the level of a setting command is ignored and
a query returns the level resulting from the total multitone signal
level.
For method selection see
The total level of the multitone signal must not exceed 1 FS.
Range:
0 FS to 1 FS
*RST:
0.05 FS
Default unit: FS
Example:
See
Firmware/Software:
V3.2.10
Manual operation:
See
SOURce:AUDio:GEN<i>:ANALog:MULTitone:STARt
<StartIndex>
SOURce:AUDio:GEN<i>:DIGital:MULTitone:STARt
<StartIndex>
Specifies the first tone of the multitone list that shall be used. The value must be equal
or less than the configured stop index.
Parameters:
<StartIndex>
Range:
1 to 20
*RST:
1
Example:
See
Firmware/Software:
V3.2.10
Manual operation:
See
SOURce:AUDio:GEN<i>:ANALog:MULTitone:STOP
<StopIndex>
SOURce:AUDio:GEN<i>:DIGital:MULTitone:STOP
<StopIndex>
Specifies the last tone of the multitone list that shall be used. The value must be equal
or greater than the configured start index.
Parameters:
<StopIndex>
Range:
1 to 20
*RST:
20
Example:
See
Firmware/Software:
V3.2.10
Manual operation:
See