GUI Reference
R&S
®
CMW-B400B/-B405A
42
User Manual 1176.8116.02 ─ 04
Fig. 4-13: Multitone generator settings
Crest Factor
Configures the crest factor of the generated multitone signal.
"Max"
All test tones of the multitone signal are generated with an initial
phase of zero degrees, resulting in the maximum possible crest fac-
tor.
"Low"
The individual test tones are generated with different initial phases,
resulting in a low crest factor.
Remote command:
SOURce:AUDio:GEN<i>:ANALog:MULTitone:CRESt
SOURce:AUDio:GEN<i>:DIGital:MULTitone:CRESt
Level Handling
You can either specify the total signal level or an individual level for each tone.
●
Total level:
Select "Use total level" and specify the "Total Level" (effective voltage or full-scale
value).
As a result all tones have the same level, displayed for information in the list. The
sum of all individual levels equals the total level.
●
Individual levels:
Select "Use separate levels" and specify the level of each tone in the list (effective
voltage or full-scale value).
The resulting "Total Level" is displayed for information.
Remote command:
SOURce:AUDio:GEN<i>:ANALog:MULTitone:LSELection
SOURce:AUDio:GEN<i>:ANALog:MULTitone:TLEVel
SOURce:AUDio:GEN<i>:DIGital:MULTitone:LSELection
SOURce:AUDio:GEN<i>:DIGital:MULTitone:TLEVel
Audio Generator Settings
135-1095-0799