Instrument Functions
R&S
®
NGP800
94
User Manual 5601.5610.02 ─ 04
Select "Yes" to restore factory adjustment.
3. To proceed channel adjustment, select "User Adjustment" in
The R&S
NGP800 displays the "ADJUSTMENT" wizard to guide the channel
adjustment procedures.
Figure 6-47: Channel adjustment wizard
4. Depending on the types of adjustment (voltage or current), setup the instruments
illustrated in
5. Follow the on-screen instructions displayed in the
.
The R&S
NGP800 applies a low voltage/current followed by a high voltage/current
for voltage/current adjustment. Key in the measured value from DMM using the on-
screen keypad. See
Figure 6-48: Channel adjustment procedure
6. Confirm the entry with
.
7. Leave the setup connections as open.
Select "Proceed" to start the voltage adjustment automatically.
8. Follow the on-screen instructions for current adjustment.
Adjustment
www.allice.de
Allice Messtechnik GmbH