Preparing for use
R&S
®
ENV216
15
User Manual 5201.6693.14 ─ 05
4
Preparing for use
Before the R&S
ENV216 is set up for operation, read the
4.1
Setting up the R&S
ENV216
Putting the R&S
ENV216 into operation
Execute the following steps in the order as described. Additional important information
is provided in separate chapters that are linked within each step.
1. Place the R&S
ENV216 on a non-flammable surface (e.g. metal plate, marble
slab).
2. Remove all surrounding items that could obstruct the ventilation openings of the
R&S
ENV216.
3. Establish an auxiliary
of the R&S
ENV216 that is
independent from the mains supply.
4. Check if the voltage of the mains supply corresponds to the
ENV216. Otherwise an external power supply can be necessary.
5.
DANGER!
Risk of electric shock. Before you connect the V-network to the power
source, you have to connect the R&S
ENV216 to a protective ground terminal. An
unearthed R&S
ENV216 is live. Touching a live electrical device causes serious
personal injury, or even death.
The protective ground connection must remain until you have disconnected the V-
network from the power supply.
Connect the R&S
6. Connect the
ENV216.
7. Connect the
to the R&S
ENV216.
Establish a reference ground connection for RF noise voltage measurement.
8.
.
Taking the R&S
ENV216 out of operation
Execute the following steps in the order as described:
1.
DANGER!
Risk of electric shock. Always disconnect the R&S
ENV216 from the
power source before disconnecting it from anything else, especially the protective
ground. An unearthed R&S
ENV216 is live. Touching a live electrical device causes
serious personal injury, or even death.
The protective ground connection must remain until you have disconnected the V-
network from the power supply.
Disconnect the R&S
ENV216 from the mains supply.
Setting up the R&S
ENV216