![Rohde & Schwarz 1424.7805K02 User Manual Download Page 501](http://html1.mh-extra.com/html/rohde-and-schwarz/1424-7805k02/1424-7805k02_user-manual_1477791501.webp)
Troubleshooting
R&S
®
NRX
501
User Manual 1178.5566.02 ─ 09
3004 - version <no.> available for sensor check source (R&S NRX-B1)
A newer version of the sensor check source (R&S NRX-B1) is available.
Solution: Update the sensor check source (R&S NRX-B1) as described in
15.2.1.4
License key notifications
Number range: 4000 to 4999.
For all notifications not described here, collect information for technical support. See
Chapter 15.4, "Collecting information for technical support"
on page 502. Described
notifications:
4001 - license key management warning
....................................................................501
4002 - remote language not allowed, NRX-K301 option missing
................................501
4001 - license key management warning
Usually caused by:
●
Problems with the system time. See
●
Wrong license key for a software option.
See
4002 - remote language not allowed, NRX-K301 option missing
For the feature you want to use, the NRX KS emulation mode (R&S NRX-K301) option
is required. The option is not installed on your R&S NRX.
Solution: Purchase the NRX KS emulation mode (R&S NRX-K301) option and install it.
See also
Chapter 11, "Option management"
15.2.1.5
Queue handling notifications
Number range starts from 9000.
9001 - static error queue overflow
The queue of events has reached its maximum number of 1000 notices, warnings and
errors. No more entries are created.
Solution: Solve the errors before continuing.
15.3
Performing tests
Using the graphical user interface, you can test the following:
●
User interface of the R&S NRX, see
●
Connected power sensors, see
In remote control, use the commands described in
Chapter 13.12, "Running selftests"
Performing tests