Remote Control - SCPI
R&S
®
QAR
84
User Manual 1178.7200.02 ─ 09
6.2
Remote Commands
.............................................................................................86
..............................................................................................87
..........................................................................................87
...........................................................................................96
6.2.1
Data Management
.......................................................................................84
................................................................................... 84
...................................................................................85
MEASurement:LOAD
<Directory>
This command restores a set of measurement results.
Setting parameters:
<Directory>
String that contains the location of the results.
The location is always a complete directory, because measure-
ment results are made up out of several files.
Example:
//Restore measurement results
MEAS:LOAD 'c:\qar\results\measurement'
Usage:
Setting only
MEASurement:RESult:DIRectory
<Directory>
This command defines the folder in which the measurement results are stored.
Parameters:
<Directory>
String containing the path to the lcoation in which results should
be stored.
*RST:
'c:\temp'
Example:
//Define result folder
MEAS:RES:DIR 'c:\results'
MEASurement:RESult:SAVE:AUTO
<State>
This command turns automatic saving of the measurement results on and off.
Parameters:
<State>
ON | 1
Automatically saves the results after the measurement is done.
Remote Commands