Remote Control - SCPI
R&S
®
QAR
85
User Manual 1178.7200.02 ─ 09
OFF | 0
Results are not saved automatically.
*RST:
ON
Example:
//Turn on automatic saving of results
MEAS:RES:SAVE:AUTO ON
MEASurement:RESult:SAVE:WHAT
<Data>
This command selects which data to store when saving.
Parameters:
<Data>
0
Stores statistics and pictures.
1
Stores statistics, pictures and data.
2
Stores statistics, pictures, data and volume.
*RST:
1
Example:
//Select data to be stored
MEAS:RES:SAVE:WHAT 0
MEASurement:SAVE
<Description>
This command saves the current measurement results.
Prerequisites for this command
●
Measurement results must be available.
●
Define the location in which the results should be stored (
Setting parameters:
<Description>
String that is the basis of the folder name in which the R&S
QAR
saves the results.
In addition to the folder description, the R&S
QAR adds a time
stamp to the folder name. For example, a description
NewResults
would create a folder
20180101-125959_NewResults
.
Example:
//Save measurement results
MEAS:STAR
MEAS:RES:DIR 'c:\results'
MEAS:SAVE 'NewResults'
Usage:
Setting only