Remote Commands
R&S
®
Spectrum Rider FPH
382
User Manual 1321.1011.02 ─ 11
The suffix at ALTernate selects one of the 11 alternate channels.
Example:
CALC:LIM:ACP:ALT2:RES?
Queries the limit check results for the second alternate chan-
nels. The command would return, e.g.
FAILED,FAILED
Upper
and lower alternate channels violate a limit.
Usage:
Query only
ACLR Result Query
The following commands query results specific to the adjacent channel power mea-
surement.
List of commands
CALCulate<1...2>:MARKer<1...6>:FUNCtion:TXPower:RESult?
......................................... 382
CALCulate<1...2>:MARKer<1...6>:FUNCtion:TXPower:RESult?
This command queries the total TX channel power. Use
MARKer<1...6>:FUNCtion:POWer:RESult?
on page 364 to query the power of
each TX channel.
The suffix at CALCulate and MARKer are irrelevant for this command.
Parameters:
<Return values>
<Power>
The power is either the peak power or the channel power,
depending on the selected standard.
Example:
CALC:MARK:FUNC:TXP:RES?
Queries the total channel power over all TX channels.
Usage:
Query only
Measuring the Harmonic Distortion
The following commands configure Harmonic Distortion measurements.. To perform
the actual measurement, use the commands described in section
"Performing and Triggering Measurements"
List of commands
CALCulate<1...2>:MARKer<1...6>:FUNCtion:HARMonics[:STATe]
...................................... 382
CALCulate<1...2>:MARKer<1...6>:FUNCtion:HARMonics:DISTortion?
................................ 383
CALCulate<1...2>:MARKer<1...6>:FUNCtion:HARMonics:LIST?
.........................................383
CALCulate<1...2>:MARKer<1...6>:FUNCtion:HARMonics:NHARmonics
..............................384
CALCulate<1...2>:MARKer<1...6>:FUNCtion:HARMonics:PRESet
......................................384
CALCulate<1...2>:MARKer<1...6>:FUNCtion:HARMonics[:STATe]
<State>
This command turns the harmonic distortion measurement on and off.
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH