Remote Commands
R&S
®
Spectrum Rider FPH
364
User Manual 1321.1011.02 ─ 11
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer:RESult?
<arg0>
This command queries the results of power measurements.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end before reading out the result. This is only possible in single sweep mode.
Before you can use this command, you have to select the power measurement with
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer:SELect
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer[:STATe]
This command is a query and therefore has no *RST value.
The numeric suffix at CALculate and MARKer are irrelevant for this command.
Suffix:
<1...2>
.
1...2
<1...6>
1...6
Parameters:
<arg0>
MCACpower | ACPower | CPOWer | EIRPpower | OBANdwidth |
OBWidth | TDMA
MCACpower | ACPower
Returns the results for ACLR measurements.
The number of return values depends on the number of TX and
adjacent channel. The order of return values is:
power of the transmission channel.
power of the lower adjacent channel.
power of the upper adjacent channel.
power of the lower alternate channel 1.
power of the upper alternate channel 2.
etc.
The unit of the return values depends on the scaling of the verti-
cal axis:
logarithmic scaling returns the power in the currently selected
unit (see
linear scaling returns the power in W.
CPOwer
Returns the results for channel power measurements.
The return value is the power of the channel. The unit depends
on the scaling of the vertical axis:
logarithmic scaling returns the power in the currently selected
unit (see
linear scaling returns the power in W.
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH