![Rohde & Schwarz 1173.8709K02 User Manual Download Page 100](http://html1.mh-extra.com/html/rohde-and-schwarz/1173-8709k02/1173-8709k02_user-manual_1477655100.webp)
Remote control basics
R&S
®
NRPxxTWG
100
User Manual 1178.8371.02 ─ 08
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
0
0
0
Power summary
0
0
0
0
Calibration summary
POST failure
0
0
0
0
0
0
+
Figure 9-3: Questionable status register
Querying the register:
●
STATus:QUEStionable:CONDition?
●
Table 9-5: Used questionable status bits and their meaning
Bit no.
Short description
Bit is set if
3
Power summary
Questionable power status register
8
Calibration summary
Questionable calibration status register
9
POST failure
Built-in test of the R&S NRPxxTWG that is carried out auto-
matically upon power-up has generated an error.
9.2.3.1
Questionable power status register
Contains information whether the measured power values are questionable.
Status reporting system