Typical Power Consumption: 2.0kW at 230V ac input
(400s pulse width and 50% duty cycle)
Power Factor:
>0.9
External Dimensions (mm):
371(W) x 270(D) x 160(H)
Weight:
13.5kg
(15.0kg with mounting bracket 105-0077-00)
(15.3kg with mounting bracket 105-0076-00)
Figure 2.2-3
External Dimensions for the DC Power Supply
– show with bracket 105-0077-00
(note: all dimensions in millimetres)
f) Coolant Requirements
Heat Load:
2000 Watts max.
Coolant flow rate:
4 litres/min (minimum)
Coolant temperature:
19 – 25
o
C (± 1
o
C from coolant setting)
Coolant pressure:
> 2.2 bar (32 psi)
< 6.0 bar (88 psi)
In line filter:
100µm
Connections:
8mm OD tubing
or barb fitting for 3/8” ID tubing
For more detailed information on the Cooling Water Specification
see section 2.3 below.
g) Environmental Specification
Ambient temperature:
+5 to +40
o
C (+41 to +104
o
F)
Storage temperature*:
-10 to +70
o
C (+14 to +158
o
F)
Humidity:
10 to 85% relative humidity
(non-condensing)
Altitude:
< 3000m