Care must be taken to avoid getting any dirt or dust on any of the optical
components.
2. Fit the DCA to the rear of the
magnifi cation block and tighten the
securing knob to hold it in place.
3. Re-fi t the eyepiece assembly to the
rear of the DCA and secure it by
tightening the securing knob.
4. Rotate the swing arm to either side of the unit,
and with a suitable Philips screwdriver remove
the cable channel cover plate.
5. Remove the cover to the USB hub on the Slit
Lamp base by undoing the small Phillips head
screw and lifting off the cover.
2.3 CONTRAINDICATION
There is no restriction to patient population this device can be used with other than those
outlined in the contra-indications stated below.
Slit lamps can produce discomfort in some photophobic patients due to the high illumination.
Additionally patients must be co-operative and capable of sitting upright for the duration of
the examination, therefore the technique may not be suitable for patients who are unable to sit
upright for long periods of time or those with limited neck and back mobility.
3. CLEANING AND DISINFECTION INSTRUCTIONS
Before any cleaning of the instrument or the base unit, ensure the power
lead is disconnected.
Only manual non-immersion cleaning as described should be used for this instrument. Do not
autoclave or immerse in cleaning fl uids. Always disconnect power supply from source before
cleaning.
1. Wipe the external surface with a clean absorbent, non-shedding cloth dampened with
de-ionised water / detergent solution (2% detergent by volume) or water / isopropyl
alcohol solution (70% IPA by volume). Avoid optical surfaces.
2. Ensure that excess solution does not enter the instrument. Use caution to ensure cloth is
not saturated with solution.
3. Surfaces must be carefully hand-dried using a clean non-shedding cloth.
4. Safely dispose of used cleaning materials.
4. INSTALLATION OF DIGITAL UNIT
For ‘Digital Ready’ variants of the Rodenstock Slit Lamp a Digital Camera Assembly (DCA) must
be purchased separately, contact Rodenstock or your local distributor for details.
Set up the Slit Lamp as directed in the Instructions For Use (IFU) EP59-51519 or EP59-70031,
Contact your distributor if you are unable to locate a copy of these Instructions.
1. Install the DCA – carefully remove the
eyepiece assembly from the
magnifi cation block by unscrewing the
securing knob at the same time as
supporting the eyepieces. The eyepieces
assembly is a dovetail type fi t to the
magnifi cation block.
Securing knob