04906900001  

INSTALLATION-MANUAL 

cobas c

 111

 

Last printed: 7 January 2010 

Page 1 of 43 

 

R

 

 
 
 
 
 
 

 

 
 
 

Installation Manual 

Version 3.0 

29-Sep-2009 

 

 
 

Summary of Contents for Cobas C111

Page 1: ...04906900001 INSTALLATION MANUAL cobas c 111 Last printed 7 January 2010 Page 1 of 43 R Installation Manual Version 3 0 29 Sep 2009...

Page 2: ...cobas c 111 Installation Manual 04906900001 INSTALLATION MANUAL cobas c 111 Version 3 0 Page 2 of 43 This page has been left empty intentionally...

Page 3: ...ation of switches and connectors 21 Main switch 21 Power cable 21 Location of fuses 21 Fluid Connections 21 USB connection 21 RS232 Serial connectors 22 6 Start up of the Instrument 23 6 1 Instrument...

Page 4: ...0001 Published by Roche Diagnostics a division of F Hoffmann La Roche 4070 Basel Switzerland Questions or comments regarding the contents of this Manual can be directed to the address below or to your...

Page 5: ...gnostics Ltd Forrenstrasse CH 6343 Rotkreuz Switzerland 1 4 Intended use The cobas c 111 instrument is a continuous random access analyzer intended for the in vitro determination of clinical chemistry...

Page 6: ...bol listing Caution Refer to accompanying documents Manufacturer of device These symbols are provided on the type plate of the instrument Electrical and electronic equipment marked with this symbol ar...

Page 7: ...lowing signs and their meanings Indicates a direct danger that if not avoided may result in death or serious injury Indicates a possibly hazardous situation that if not avoided may result in death or...

Page 8: ...vide any anti virus Software Therefore it is essential to take precautions to ensure that any software loaded onto the system is virus free Do not load any software that is not approved by Roche Diagn...

Page 9: ...local regulations Please refer to your local safety regulations regarding handling sample material Waste Danger from infectious waste material can cause severe illness Avoid direct contact with waste...

Page 10: ...aining material of human origin must be treated as potentially infectious Always wear protective gloves when handling this material Do not touch parts of the analyzer other than those specified The wh...

Page 11: ...lume per test from 2 15 l Manual and or computer interfaced test ordering Cooled storage area for Roche reagents 27 reagent bottles per reagent disk and up to 8 reagent disks can be handled Using reag...

Page 12: ...cobas c 111 Installation Manual 04906900001 INSTALLATION MANUAL cobas c 111 Version 3 0 Page 12 of 43 4 2 Instrument internal layout The major components of the analyzer are shown below in fig 1...

Page 13: ...D VGA 320 x 240 pixels passive matrix touch screen Printer Internal thermal printer 112 mm paper width Samples Sample handling Direct tube placement manually by user Incubation time Max 12 min 40 cycl...

Page 14: ...e con This unit must be connected to a grounded mains outlet nected to a grounded mains outlet nected to a grounded mains outlet nected to a grounded mains outlet Insulation coordination Category II I...

Page 15: ...Union EU Directive 98 79 EC in vitro medical device FCC This equipment complies with the requirement in part 15 of FCC rules for a class A computing device Operation of this equipment in a residential...

Page 16: ...f this document 5 1 1 Packing list The actual version of the analyzer as delivered together with a list of the accessories for your configuration are detailed in the accompanying packing list That lis...

Page 17: ...he warranty on the analyzer The original shipping containers must be transferred unopened to the installation site Unpack the instrument and check that all items mentioned on the packing list are pres...

Page 18: ...ON MANUAL cobas c 111 Version 3 0 Page 18 of 43 2 Remove the standard accessories 3 Lift the carton off the pallet to get access to the instrument 4 Remove the upper cardboard cover 5 Remove the plast...

Page 19: ...111 on flat and stable surface Notice Keep the packing pieces in a safe place to use again whenever the instru ment has to be transported any significant distance 5 3 2 Unpacking procedure for instrum...

Page 20: ...on off the pallet to get access to the instrument 4 Remove the upper cardboard cover 5 Remove the plastic transport protection covering the instrument 6 Lift the cobas c 111 instrument from the pallet...

Page 21: ...the rear of the instrument Important Only use an approved power supply cord in good condition Do not touch mains cables if your hands are wet Do not attempt to connect or disconnect the mains cables...

Page 22: ...22 of 43 RS232 Serial connectors There are 2 serial interface connectors intended for use with a Barcode scanner and Host interface respectively Please refer to the cobas c 111 Operator Manual and the...

Page 23: ...Operator s Manual Notice If there is a Reagent Disk inside but is not detected by the software then check for correct identification by the flags Exactly one of these flags needs to be broken away for...

Page 24: ...p watch the valves to operate correctly 18 Check if the peristaltic pump cover is closed 19 Use the maintenance action Prime ISE calibrator and reference to prime the tubing 20 Use the maintenance act...

Page 25: ...ation report form Any factory set calibration data found on the instrument may be outdated and therefore cause wrong results Cross check the calibration results and the application producing correct r...

Page 26: ...ide of the instrument Place a new bottle of cleaner fill the Water Container with de ionized water CLSI Grade II and check that the Waste Container is empty 7 As next step Connect the Cleaner Protecti...

Page 27: ...ter when delivered eventually an error message Air detected in the fluid system will appear Ignore it until you performed step 6 Otherwise you have to check the fluid tubing for leaks 5 Check the Reag...

Page 28: ...tion result for the CHECK test that is stored in the database can not be deleted as it is always used as reference 10 Reset Maintenance Counter Please choose Utilities Diagnostics Analyzer Installatio...

Page 29: ...cover 5 Remove the screws on the left and right side that holds the rear service flap and fold it backwards 6 Remove the complete ISE Cover and left side panel of the ISE unit 7 Route the flat ribbon...

Page 30: ...INSTALLATION MANUAL cobas c 111 Version 3 0 Page 30 of 43 9 Connect the flat ribbon cable to the PCB ISE Controller J2 10 Remove the power box guard plate ISE Module fixation screws Flat ribbon cable...

Page 31: ...0001 INSTALLATION MANUAL cobas c 111 Version 3 0 Page 31 of 43 11 Connect the ground strap using the provided screw and serrated lock washer 12 Connect the waste tube to the corresponding plug ISE gro...

Page 32: ...ap into normal position and fix it with the 2 screws 16 Put the right side panel back into place 17 If not yet mounted fix the air tube to the mixtower socket Mount the sample inlet tube to the socket...

Page 33: ...to the rear connector of the internal waste tank 20 Finally mount the waste tube to ISE tower mount the tower itself and lock the ring Apply the overflow adapter at the end 21 Connect the ISE power s...

Page 34: ...9 1 Drain the instrument with ISE of water and cleaner 1 Perform the Diagnostic function Drain ISE Unit follow its instructions 2 Store the valve caps and clamps and the electrodes in separate bags a...

Page 35: ...age 35 of 43 9 3 Packing procedure for instrument with ISE Notice Use only the original packing pieces to prevent any damage of the instrument Notice The packing instructions are also shown on a label...

Page 36: ...strument and place it into the cellular material Caution The instrument weighs around 38 kg 80 lbs 3 Apply the protection plastic foil and the plastic bag over the transfer head 4 Set up the left and...

Page 37: ...L cobas c 111 Version 3 0 Page 37 of 43 5 Impose on the surrounding carton 6 Put on the cardboard cover Packed in surrounding carton Transfer head protection Make sure to have the transfer head protec...

Page 38: ...ttles into the other box 8 If the TIP N TELL indicator applied on the cardboard cover iniside box and outside are still ok leave them otherwise cancel the corresponding one with a pen 9 If possible ap...

Page 39: ...es to prevent any damage of the instrument Notice The packing instructions are also shown on a label outside of the box 1 Start with preparing the palette 2 Lift the cobas c 111 instrument and place i...

Page 40: ...sfer head Set up the left and right protection pieces just turn the one on the left side by 180 degrees 4 Impose on the surrounding carton 5 Put on the cardboard cover Instrument with protection plast...

Page 41: ...fluid bottles into the other box 7 If the TIP N TELL applied on the cardboard cover inside box and outside are still ok leave them otherwise cancel the corresponding one with a pen 8 If possible appl...

Page 42: ...d out wheeled bin symbol are covered by the European directive 2002 96 EC WEEE These items must be disposed of via designated collection facilities appointed by government or local authorities WEEE Di...

Page 43: ...Date Country City Responsible FSE Main Power Supply Voltage Frequency Subject Not OK OK Comments Package Accessories Chassis Housing Robotic Transfer Fluid Pipetting Internal Printer Analyzer Assembly...