function. Follow the instructions given by the software, when and how to place
these tubes.
8.
Check the correct setting of time and date.
9.
As a general function test run a single CHECK test first. For this purpose Import
and Install the CHECK test application and order a single test only, like you
would order any other test. Check that the result is within the range printed on
the CHECK test solution bottle. Later you may disable the CHECK test.
Please note:
The calibration result for the CHECK test that is stored in the database can not
be deleted, as it is always used as reference.
10.
Reset Maintenance Counter: Please choose “Utilities -> Diagnostics -> Analyzer -
> Installation Completed”
11.
Prepare the requested reagent bottles and insert them into the instrument.
12.
Prepare the calibrators and QC material accordingly and calibrate all tests.
13.
Check the test performance by running controls.
14.
Complete the installation report form.
Any factory-set calibration data found on the instrument may be outdated and
therefore cause wrong results! Cross-check the calibration results and the
application producing correct results by running qualified control material. Refer
to the
cobas c
111 Operator’s Manual for more handling information.
Warning